Programming

 

 

 

 

Capacitance Compensation

 

 

Table 4-13

Capacitance Compensation Program Example

 

 

 

 

 

Imports Agilent.TMFramework

 

 

Imports Agilent.TMFramework.DataAnalysis

 

 

Imports Agilent.TMFramework.DataVisualization

 

Imports Agilent.TMFramework.InstrumentIO

 

 

Imports Agilent.TMFramework.InstrumentDriverInterop

 

Imports Agilent.TMFramework.InstrumentDriverInterop.Design

 

Imports Agilent.TMFramework.InstrumentDriverInterop.VxipnpWrappers

 

Module Module1

 

’9

Sub Main()

 

Dim Agb2200 As Agb220xa = New Agb220xa("GPIB0::22::INSTR", True, True)

 

Agb2200.Reset()

 

’12

Dim f_com as String = "C:\temp\my_env_1.txt"

Agb2200.SelectCompenFile(f_com)

 

 

Dim freq As Double = 1000000

’measurement frequency: 1 (MHz)

15

Dim data_c As Double = 0.0000000001

’C measured by 4284A: 100 (pF)

 

Dim data_g As Double = 0.0005

’G measured by 4284A: 0.5 (mS)

 

Dim res_c As Double

 

 

Dim res_g As Double

 

 

Agb2200.CompenC(freq, data_c, data_g, res_c, res_g)

’20

Dim result As String = "C = " & res_c * 1000000000000.0 & " (pF)"

’22

result = result & Chr(10) & "G = " & res_g * 1000 & " (mS)" & Chr(10)

 

Console.WriteLine(result)

 

 

Agb2200.Close()

 

 

End Sub

 

 

 

End Module

 

 

 

Line

 

Description

 

 

 

 

 

1 to 11

 

The above example is for the B2200 of the GPIB address 22 on the interface GPIB0.

 

 

 

“GPIB0” is the VISA name. Confirm your GPIB settings, and set them properly.

 

 

 

 

 

12 to 13

 

The lines specify the compensation data file. The file name must specify your

 

 

 

compensation data file.

 

 

 

 

 

15 to 20

 

Compensates the data measured by the Agilent 4284A. In this example, the measurement

 

 

frequency is 1 MHz, the capacitance data is 100 pF, and the conductance data is 0.5 mS.

 

 

Change freq, data_c, data_g values for your measurement results.

 

 

 

 

22 to 24

 

Displays the compensation result data on the console window. Record the values as C1r and

 

 

G1r, or C2r and G2r.

 

 

 

 

 

 

 

4-28

Agilent B2200 User’s Guide, Edition 2

Page 126
Image 126