SCPI Command Reference

Command Summary

DIAG subsystem :DIAGnostic subsystem has commands for executing the self-test function. For more info, see “Selftest Menu” on page 3-34.

The :DIAGnostic subsystem commands ignore the B2200 channel configuration mode. For :DIAG commands that require a card number, you specify 1, 2, 3, 4, or ALL.

Command

Description

 

 

:DIAG:TEST:CARD:CLE card_number

Clears relay test result (pass/fail) of the specified card.

 

card_number: 1, 2, 3, 4, or ALL

:DIAG:TEST:CARD[:EXEC]? card_number

Executes relay test, then returns result: 1 (fail card exists), 0

 

(pass).

 

card_number: 1, 2, 3, 4, or ALL

:DIAG:TEST:CARD:STAT? card_number

Returns most recent relay test result: 1 (fail), 0 (pass), -1(not

 

tested).

 

card_number: 1, 2, 3, 4

:DIAG:TEST:FRAM:CLE item

Clears specified test result.

 

item: CONT (controller test), FPAN (front panel interface test),

 

LED, PEN, or BEEP

 

 

:DIAG:TEST:FRAM[:EXEC]? item

Executes specified test, then returns test result: 1 (fail), 0

 

(pass).

 

item: CONT (controller test), FPAN (front panel interface test),

 

LED, PEN, or BEEP

 

 

:DIAG:TEST:FRAM:STAT? item

Returns most recent test result of the specified test: 1 (fail), 0

 

(pass), -1(not tested).

 

item: CONT (controller test), FPAN (front panel interface test),

 

LED, PEN, or BEEP

 

 