SCPI Command Reference [:ROUTe]:COUPle:PORT:DETect

[:ROUTe]:COUPle:PORT:DETect

This command detects the input ports connected to the Kelvin cable, and assigns them as the input couple ports that will be used for the Kelvin connection. The input couple port setting is effective for the all cards. This command overwrites the previous couple port setting.

The couple mode is set by “[:ROUTe]:COUPle[:STATe]”. The couple port setting will be cleared and updated by “[:ROUTe]:COUPle:PORT”.

The input port numbers of the couple ports can be got by [:ROUTe]:COUPle:PORT?.

After this command is executed, the all relay settings will be the same as after the :ROUT:OPEN:CARD ALL command execution.

Syntax[:ROUTe]:COUPle:PORT:DETect

Example OUTPUT @Agb2200;":ROUT:COUP:PORT:DET"

5-38

Agilent B2200 User’s Guide, Edition 2

Page 164