Front Panel Operation

Operation

To Use Ground Mode

Ground mode cannot be set to ON when the bias mode is ON.

1.Change the ground-enabled output ports.

a.Press the Shift key and the Open All key.

b.Move the cursor to SET, then press the Enter key.

c.Move the cursor to ENABLE, then press the Enter key.

The LCD displays one of the ground-disabled output port numbers which can be changed to the ground-enabled. Or if there is no ground-disabled output ports, the LCD displays Channel: No Channel. Press the arrow key to select the output port to be changed to the ground-enabled, then press the Enter key.

d.Move the cursor to DISABLE, then press the Enter key.

The LCD displays one of the ground-enabled output port numbers which can be changed to the ground-disabled. Or if there is no ground-enabled output port, the LCD displays Channel: No Channel. Press the arrow key to select the port to be changed to the ground-disabled, then press the Enter key.

e.Repeat c and d until the ground-enabled port setup is completed. After that, press the Exit key more than once to display the status and setup conditions.

2.Press the Shift key and the Couple Mode key to set the ground mode ON. Performing this operation again sets the mode OFF.

3.Change the ground port (default setting: 12).

a.When the ground mode is ON, press the Port Function key. The blink-cursor appears on the first line of the LCD.

b.Press the right/left arrow key to move the cursor on the input port number (01 to 14) that will be set to the input ground port.

c.Press the up/down key to display G.

4.Change the ground-enabled input ports (multiple ports can be set). Repeat a and

bfor the all ground-enabled input ports, then press the Enter key.

a.Press the right/left arrow key to move the cursor on the input port number (01 to 08) that will be set to the ground-enabled input port.

b.Press the up/down key to display -.

Open the ground-enabled input ports to prevent the instrument from damage.

3-8

Agilent B2200 User’s Guide, Edition 2

Page 68
Image 68
Agilent Technologies B2201A manual To Use Ground Mode