Installation

Measurement Cable Length

Table 2-8

Guard Capacitances of B2200 Measurement Environment (Typical)

 

 

 

 

 

Connection

Agilent Model/Part No.

Cable

Guard

Length

Capacitance

 

 

 

 

 

 

 

 

from SMU to B2200

16494A-003 (for non-Kelvin)

80 cm

75 pF

inputs

 

 

 

 

 

16494A-001 (for non-Kelvin)

1.5 m

130 pF

 

 

 

 

 

 

 

 

 

16494B-003 (for Kelvin, 4142B)

80 cm

90 pF

 

 

 

 

 

 

 

16493K-001 (for Kelvin)

1.5 m

150 pF

 

 

 

 

mainframe and

B2210A (only 1 card is installed)

145 pF

modules

 

 

 

 

 

B2210A (more 1 card)

8 pF/extra card

 

 

 

 

 

 

 

 

 

B2211A (only 1 card is installed)

145 pF

 

 

 

 

 

 

 

B2211A (more 1 card)

8 pF/extra card

 

 

 

 

from B2200 outputs

16494A-001 (for non-Kelvin)

1.5 m

130 pF

to connector plate or

 

 

 

 

 

16494A-002 (for non-Kelvin)

3 m

240 pF

probe card interface

 

 

 

 

 

16494A-003 (for non-Kelvin)

4 m

340 pF

 

 

 

 

 

 

 

16494B/C-001 (for Kelvin)

1.5 m

140 pF

 

 

 

 

 

 

 

16494B/C-002 (for Kelvin)

3 m

260 pF

 

 

 

 

 

 

 

16494C-003 (for Kelvin)

4 m

340 pF

 

 

 

 

from connector

Agilent Part No. 8121-1191

X m

130 pF per m

plate to DUT

 

 

 

 

 

 

 

probe card interface

B2220A

70 pF

 

 

 

 

 

2-24

Agilent B2200 User’s Guide, Edition 2

Page 58
Image 58