Agilent Technologies B2201A Cannot use same port for Unused and Couple, Bad input port number

Models: B2201A

1 234
Download 234 pages 49.02 Kb
Page 234
Image 234

Error Messages

B2200 Specific Error Messages

Error

Error Message and Description

Number

 

 

 

3026

Cannot use same port for Unused and Couple

 

The couple mode and the ground mode cannot be used in parallel

 

when a couple port and the ground enabled input port (unused port)

 

have been assigned to the same input port.

 

 

3027

Cannot use Unused Port during Auto Ground Mode ON

 

The ground enabled input port (unused port) must be opened during

 

the ground mode is ON.

 

 

3030

Bad input port number

 

The input port number must be 1 to 14.

 

 

3031

Bad output port number

 

The output port number must be 1 to 12, 24, 36, or 48. Maximum

 

number depends on the switch module configuration.

 

 

3032

Bad setting memory number

 

The internal memory number for the setup data must be 1 to 8.

 

 

3033

EEPROM programming failure

 

Cannot change the GPIB address or update internal memory data.

 

Contact your nearest Agilent Technologies service center.

 

 

3034

EEPROM reading failure

 

Cannot read the GPIB address or internal memory data. Contact your

 

nearest Agilent Technologies service center.

 

 

3035

Setting memory data is invalid

 

Cannot read the setup data in the internal memory. The data has been

 

broken. Delete the data.

 

 

3036

Cannot load this setting data in this configuration

 

Cannot read the setup data in the internal memory due to the

 

configuration mismatch. Delete the data or leave it until restoring the

 

module configuration.

 

 

7-14

Agilent B2200 User’s Guide, Edition 2

Page 234
Image 234