Front Panel Operation

Operation

To Initialize Agilent B2200

1.Press the Shift key and the Local key. Then the LCD will display NO.

2.Press the arrow key to displays YES, then press the Enter key to initialize the Agilent B2200, or press the Exit key to cancel initialization.

To Enable Light Pen

Turn the Agilent B2200 off, and connect the light pen to the Light Pen connector at the right down corner of the front panel. After that, perform the following procedure to enable the light pen.

1. Press the Menu key. The setup menu will be displayed on the LCD.

2. Press the arrow key to move the cursor to PEN, then press the Enter key.

3. Press the arrow key to display ENABLE, then press the Enter key to enable the light pen, or press the Exit key to cancel changing the setup.

NOTE

Light Pen

 

You can control the Agilent B2200 functions and the switch connections by using

 

the front panel keys. And the Agilent B2200 allows you to use the Agilent 16443A

 

Light Pen to control the switch connections.

 

Pointing a LED on the LED matrix by using the light pen controls the setting of the

 

switch specified by the LED.

To Change Channel Configuration Mode

1.Press the Shift key and the Bias Mode key. The LCD displays the present setting (AUTO or NORMAL).

2.Press the arrow key to select the mode, then press the Enter key. The LCD displays NO.

3.Press the arrow key to display YES, then press the Enter key. Or press the Exit key to cancel this operation.

Changing the channel configuration mode resets the Agilent B2200.

3-4

Agilent B2200 User’s Guide, Edition 2

Page 64
Image 64
Agilent Technologies B2201A manual To Initialize Agilent B2200, To Enable Light Pen, To Change Channel Configuration Mode