Programming
Capacitance Compensation
To Perform Measurement and Compensation
Perform the capacitance measurement and compensation as shown below.
1.Set the Agilent 4284A measurement condition. Then the frequency must be the value (Fmeas) used when the compensation coefficients are measured.
2.Before contacting the device under test (DUT), perform the
3.Perform the compensation by using the capacitance compensation function, and record the result data (C1r and G1r).
See Table
agb220xa_selectCompenFile
Also, agb220xa_reset and agb220xa_close have been changed.
4.Contact the DUT, perform the
5.Perform the compensation and record the result data (C2r and G2r).
6.Perform the following calculation and record it as the capacitance value.
C = C2r - C1r
Table | Recording Measurement/Compensation Data |
| |||
|
|
|
|
|
|
|
|
| step | Measurement/Compensation Data | |
|
|
|
|
| |
|
|
| C (F) | G (S) | |
|
|
|
| ||
|
|
|
|
|
|
|
| 2 | (measured) | C1 = | G1 = |
|
|
|
|
|
|
|
| 3 | (compensated) | C1r = | G1r = |
|
|
|
|
|
|
|
| 4 | (measured) | C2 = | G2 = |
|
|
|
|
|
|
|
| 5 | (compensated) | C2r = | G2r = |
|
|
|
|
|
|
|
| 6 | (calculated) | C = |
|
|
|
|
|
|
|
Agilent B2200 User’s Guide, Edition 2 |