Programming

Capacitance Compensation

To Perform Measurement and Compensation

Perform the capacitance measurement and compensation as shown below.

1.Set the Agilent 4284A measurement condition. Then the frequency must be the value (Fmeas) used when the compensation coefficients are measured.

2.Before contacting the device under test (DUT), perform the Cp-G measurement in the open condition at the end of the measurement path including positioner or probe card, and record the measurement data (C1 and G1). See Table 4-12.

3.Perform the compensation by using the capacitance compensation function, and record the result data (C1r and G1r).

See Table 4-13 for the example to use the capacitance compensation function. This example uses Microsoft Visual Basic .NET and Agilent T&M Programmers Toolkit. And the Agilent B2200 controlled by the program is defined as AGB2200 at the 10th line. So, the function names are changed as shown below.

agb220xa_selectCompenFile --> Agb2200.SelectCompenFile agb220xa_compenC --> Agb2200.CompenC

Also, agb220xa_reset and agb220xa_close have been changed.

4.Contact the DUT, perform the Cp-G measurement, and record the measurement data (C2 and G2).

5.Perform the compensation and record the result data (C2r and G2r).

6.Perform the following calculation and record it as the capacitance value.

C = C2r - C1r

Table 4-12

Recording Measurement/Compensation Data

 

 

 

 

 

 

 

 

 

 

step

Measurement/Compensation Data

 

 

 

 

 

 

 

 

C (F)

G (S)

 

 

 

 

 

 

 

 

 

 

 

 

2

(measured)

C1 =

G1 =

 

 

 

 

 

 

 

 

3

(compensated)

C1r =

G1r =

 

 

 

 

 

 

 

 

4

(measured)

C2 =

G2 =

 

 

 

 

 

 

 

 

5

(compensated)

C2r =

G2r =

 

 

 

 

 

 

 

 

6

(calculated)

C =

 

 

 

 

 

 

 

Agilent B2200 User’s Guide, Edition 2