SCPI Command Reference [:ROUTe]:SYMBol:PORT

[:ROUTe]:SYMBol:PORT

In the GPIB local mode, the B2200 uses 01 to 14 (two digits numbers) to specify the input ports 1 to 14 respectively. This command specifies a symbol string and lets the B2200 use the specified string instead of the two digits number. The query returns the symbol string.

Reboot and *RST clear the symbol strings. After that, the initial symbol string (two digits numbers 01 to 14) will be set.

:SYSTem:CPON does not clear the symbol strings.

Syntax

[:ROUTe]:SYMBol:PORT port_number,'symbol_string'

 

[:ROUTe]:SYMBol:PORT? port_number

 

 

 

 

Parameter

Explanation

 

 

 

 

port_number

Input port number: 1 to 14.

 

symbol_string

Symbol string set to the specified input port. Maximum 6

 

 

characters.

 

 

Numeric characters, alphabetic characters; upper case and

 

 

lower case, space, and the following symbols are available.

 

 

! " # $ % & ’ ( ) * + , - . / > ? : ~ ; < = > ? @ [ \ ] ‘ { } ^ _

 

 

 

Query response

symbol_string <newline><^END>

Example

OUTPUT @Agb2200;":ROUT:SYMB:PORT 1,'SMU1'"

 

OUTPUT @Agb2200;":ROUT:SYMB:PORT? 1"

 

ENTER @Agb2200;A$

This example assigns the string SMU1 to the input port 1. After that, SMU1 is used instead of 01. In this example, A$ will be SMU1.

Agilent B2200 User’s Guide, Edition 2

5-43

Page 169
Image 169
Agilent Technologies B2201A manual ROUTeSYMBolPORT