Agilent Technologies B2201A manual Agb220xagroundChanListQ, Gnddisen, Gndstatus

Models: B2201A

1 234
Download 234 pages 49.02 Kb
Page 211
Image 211

VXI plug&play Driver

Driver Functions

agb220xa_groundChanList_Q

This function will query the instrument for the status, ground enabled or disabled, for the channels given in the list.

The parameter “gndchan_list” is an array of integers with each integer representing one channel. The last number of the “gndchan_list” should be “0” (numeric zero) to identify the end of the list. The maximum number of channels that can be specified by the list is 100.

The “gnd_status” parameter is an array of integers containing the return values of the query. The “gnd_status” array returned will correspond one to one with “gndchan_list” parameter.

Syntax

ViStatus _VI_FUNC agb220xa_groundChanList_Q(ViSession vi,

 

ViInt16 gnd_disen, ViInt32_VI_FAR gndchan_list[ ],

 

ViInt32_VI_FAR gnd_status[ ] );

Parameters

vi

Instrument handle returned from agb220xa_init( ).

 

gnd_disen

Query type.

 

 

0 : checks if the port is the ground enabled.

 

 

1 : checks if the port is the ground disabled.

 

gndchan_list[ ]

Channel numbers to check the status. 5 digits integer. ABCDE.

 

 

where A: card number, BC: input port number, DE: output port

 

 

number. Top zero(s) can be ignored. For example, if A=0,

 

 

BC=01, and DE=01, channel number should be 101 instead of

 

 

00101.

 

gnd_status[ ]

Status of the channels given in the gndchan_list. Returned value

depends on the setting of gnd_disen as shown below:

when gnd_disen=0, 0 means ground disabled, 1 means enabled. when gnd_disen=1, 0 means ground enabled, 1 means disabled.

Agilent B2200 User’s Guide, Edition 2

6-25

Page 211
Image 211
Agilent Technologies B2201A manual Agb220xagroundChanListQ, Gnddisen, Gndstatus