Installation

Output Connections

NOTE

To Check Interlock Circuit

 

If you use the 4155/4156, you can easily check the interlock circuit as follows:

 

1.

Connect the Intlk connector of the 4155/4156 to your interlock circuit.

 

2.

Press System front-panel key, then select CALIB/DIAG primary softkey to

 

 

display the SYSTEM: SELF-CALIBRATION/DIAGNOSTICS page.

 

3.

In the CALIB/DIAG field, select DIAG secondary softkey.

 

4.

In the CATEGORY field, select I/O PERIPH secondary softkey.

 

5.

Move pointer to the 403 (INT.) Interlock LED field.

 

6.

Select EXECUTE secondary softkey.

 

7.

Confirm the following:

 

 

• LED turns on within 1 sec from when interlock circuit is shorted.

 

 

• LED turns off within 1 sec from when interlock circuit is open.

 

To stop the interlock test, select STOP secondary softkey.

 

 

 

To Mount Connectors

This section provides the information needed to mount connectors for the Agilent B2200 output cable connections on the shielding panel. See this section if you do not use the connector plate.

1. Get the appropriate parts for your situation. Refer to Table 2-5.

2. Make holes and mount the connectors. Refer to Table 2-6.

For kelvin connections, use the Kelvin cables. The Kelvin cable requires the kelvin triaxial connector which has two connector holes and three screw holes.

3. Make an interlock circuit as shown in “To Make Interlock Circuit” on page 2-15.

4. Make cable connections to the DUT interface. Refer to “To Make Connections to DUT Interface” on page 2-13.