NOTE

Installation

Self-Test

Self-Test

To confirm the specifications

The self-test and diagnostics checks the operation of the mainframe and the modules. However they cannot confirm if the Agilent B2200 satisfies its specifications.

For verifying the specifications, contact your nearest Agilent Technologies Service Center. Trained service personnel will perform calibration (performance verification).

It is recommended to perform calibration once a year at least.

The following procedure performs the self-test and diagnostics.

1.Turn on the Agilent B2200.

The controller test will be performed. If the Agilent B2200 fails the test, contact your nearest Agilent Technologies Service Center.

2.Press the Shift key and the Menu key in this order. Self-test menu will appear on the LCD.

SELF_TEST

Controller test

RELAY_TEST

Relay test

KEY

Front panel interface test

BEEPER

Beeper test

LED

LED matrix test

PEN

Light pen test

GPIB

GPIB test

3.Move the cursor to the test item to perform, then press Enter key.

4.To start the test, move the cursor to EXECUTE, then press Enter key.

To display the previous test result, move the cursor to RESULT, then press Enter key.

For the details of each test, see “Selftest Menu” on page 3-34.

2-10

Agilent B2200 User’s Guide, Edition 2

Page 44
Image 44
Agilent Technologies B2201A manual Self-Test