VXI plug&play Driver

 

 

Driver Functions

 

agb220xa_biasPort

 

This function will select which input port is the bias port on the specified card. For

 

each card, you can specify the same or different bias port.

Syntax

ViStatus _VI_FUNC agb220xa_biasPort(ViSession vi, ViInt16 bport_cardno,

 

ViInt16 bias_port);

 

Parameters

vi

Instrument handle returned from agb220xa_init( ).

 

bport_cardno

Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5

 

 

(all card) in the normal configuration mode, or 0 (all card in the

 

 

auto configuration mode). For the configuration mode, see

 

 

agb220xa_func.

 

bias_port

Input port number to be set to the bias port. 1 to 14 (input port 1

 

 

to input port 14).

 

If the bias input port and a couple port have been assigned to the same input port, the

 

bias mode and the couple mode cannot be used in parallel.

 

agb220xa_biasState

 

This function controls the bias mode for the specified card. When bias mode is ON,

 

the input bias port is connected to all bias enabled output ports that are not

 

connected to any other input ports. Bias disabled output ports are never connected to

 

the input bias port when bias mode is ON.

 

The bias mode cannot be set to ON when the ground mode is ON.

Syntax

ViStatus _VI_FUNC agb220xa_biasState(ViSession vi, ViInt16 bstate_cardno,

 

ViInt16 state);

 

Parameters

vi

Instrument handle returned from agb220xa_init( ).

 

bstate_cardno

Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card 4), or 5

 

 

(all card) in the normal configuration mode, or 0 (all card in the

 

 

auto configuration mode). For the configuration mode, see

 

 

agb220xa_func.

 

state

Bias mode. 0 (OFF) or 1 (ON).

Agilent B2200 User’s Guide, Edition 2

6-9

Page 195
Image 195
Agilent Technologies B2201A manual Agb220xabiasPort, Agb220xabiasState