Saving Input/Output Labels

 

You can define labels to use for the input/output ports when you control the switch

 

in the GPIB local mode. The labels are cleared by *RST. So it is recommended to

 

save the labels into the internal memory and define a comment for the memory data.

 

You can see the comment when you load/save the memory data in the GPIB local

 

mode.

Setup:

• Channel configuration mode: Auto

 

• Label for input 1: SMU1

 

• Label for input 2: SMU2

 

• Label for input 3: SMU3

 

• Label for input 4: UNUSED

 

• Label for input 5: SMU4-F

 

• Label for input 6: SMU4-S

 

• Label for input 7: SMU5-F

 

• Label for input 8: SMU5-S

 

• Label for input 9: OPEN

 

• Label for input 10: BIAS

 

• Label for input 11: OPEN

 

• Label for input 12: GROUND

 

• Label for input 13: CMU-H

 

• Label for input 14: CMU-L

 

• Label for output 1: BULK

 

• Label for output 2: SOURCE

 

• Label for output 3: GATE

 

• Label for output 4: DRAIN

 

• Label for output 5: GROUND

 

• Comment for memory 1: Port label info