Agilent Technologies B2201A manual DIAGnosticTESTCARDCLEar, DIAGnosticTESTCARDEXECute?

Models: B2201A

1 234
Download 234 pages 49.02 Kb
Page 147
Image 147

 

 

 

SCPI Command Reference

 

 

 

:DIAGnostic:TEST:CARD:CLEar

 

 

:DIAGnostic:TEST:CARD:CLEar

 

 

This command clears the relay test result (pass/fail result) of the specified card.

Syntax

 

:DIAGnostic:TEST:CARD:CLEar card_number

 

 

 

 

 

 

Parameter

Explanation

 

 

 

 

 

 

card_number

card number: 1, 2, 3, 4, or ALL

Example

 

OUTPUT @Agb2200;":DIAG:TEST:CARD:CLE 1"

 

 

:DIAGnostic:TEST:CARD[:EXECute]?

 

 

This command executes the relay test for specified card, then returns the pass/fail

 

 

result. Before starting the relay test, open the input/output terminals (end of cable is

 

 

OK, Kelvin cable must be removed). For more information, see “RELAY_TEST”

 

 

on page 3-35.

 

 

 

This command changes all settings of card to same state as after executing

NOTE

 

 

 

:SYST:CPON command.

 

Syntax

 

:DIAGnostic:TEST:CARD[:EXECute]? card_number

 

 

 

 

 

 

Parameter

Explanation

 

 

 

 

 

 

card_number

card number: 1, 2, 3, 4, or ALL

Query response

 

test_result <newline><^END>

 

 

1: failure card exists

 

 

 

0: pass

 

Example

 

OUTPUT @Agb2200;":DIAG:TEST:CARD? ALL"

Agilent B2200 User’s Guide, Edition 2

5-21

Page 147
Image 147
Agilent Technologies B2201A manual DIAGnosticTESTCARDCLEar, DIAGnosticTESTCARDEXECute?