WARNING
CAUTION
CAUTION
Installation
Installing the B2200
To Install the Switch Module
To prevent electrical shock, turn off the mainframe and remove the power cable before starting the instruction.
Be careful about the module pins used for internal connection to the Agilent B2200. The pins can be damaged easily.
Use clean handling and
The following procedure explains the module installation and removal.
1.Turn off the Agilent B2200, then wait at least 10 seconds before you remove or install a module.
2.Loosen the screws on both the left and right edges of a blank panel or a module attached to the slot you want to install a new module.
3.Remove the blank panel or the module.
4.Align the new module with the left and right slot guide rails. Then the component side should be facing up.
5.Push the module into the slot until you feel it seat firmly into the connector at the back of the slot.
6.Screw in the screws on the left and right edges of the module.
7.Execute the
To Install the Blank Panel
To prevent thermal damage to the module, be sure that blank panels are installed in all unused slots.
If the blank panel is not installed to cover an unused slot, install the blank panel as follows:
1.Align the blank panel over the unused slot.
2.Screw in the screws on the left and right edges of the blank panel.
Agilent B2200 User’s Guide, Edition 2 |