Agilent Technologies B2201A manual Connection Sequence Relay Operation

Models: B2201A

1 234
Download 234 pages 49.02 Kb
Page 75
Image 75

 

 

 

 

 

Front Panel Operation

 

 

 

 

 

Switch Control Functions

 

Connection Sequence

 

 

 

 

This is only for modules that are set to the single connection rule. Refer to

NOTE

 

“Connection Rule” on page 3-14.

 

 

 

Connection sequence specifies the open/close sequence of the relays when changing

 

from an existing connection to a new connection.

 

You can select one of the three connection sequences shown in Table 3-1 for each

 

module installed in the Agilent B2200.

 

When an existing connection is changed to a new connection, the relays are

 

opened/closed as follows:

 

 

 

Table 3-1

Connection Sequence

 

 

 

 

 

 

 

 

 

 

 

 

Connection

 

 

Relay Operation

 

 

 

Sequence

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Break Before Make

 

1.

Disconnect previous route.

 

 

 

 

 

2.

Wait for relays to open.

 

 

 

 

 

3.

Connect new route.

 

 

 

 

 

 

 

 

 

 

Make Before Break

 

1.

Connect new route.

 

 

 

 

 

2.

Wait for relays to close.

 

 

 

 

 

3.

Disconnect previous route.

 

 

 

 

 

 

 

 

 

 

No Sequence

 

1.

Disconnect previous route.

 

 

 

 

 

2.

Connect new route.

 

 

 

 

 

 

 

 

Agilent B2200 User’s Guide, Edition 2

3-15

Page 75
Image 75
Agilent Technologies B2201A manual Connection Sequence Relay Operation