Front Panel Operation
Operation
To Change Connection Rule
1.Press the Rule key. The LCD displays the present setting (FREE or SINGLE).
2.Press the arrow key to select the connection rule, then press the Enter key. Or press the Exit key to cancel this operation.
To Change Connection Sequence
1.Press the Shift key and the Rule key. The LCD displays the present setting (Break_Before_Make, Make_Before_Break, or No_Sequence).
2.Press the arrow key to select the connection sequence, then press the Enter key. Or press the Exit key to cancel this operation.
To Control Switch Condition
If you use the light pen, you do not need the following instruction to control the switch condition.
1.If you are NOT watching the
2.Press the arrow keys to move the
3.Press the Open/Close key to open/close the switch specified by the
4.Repeat 2 and 3 until the switch setup is completed.
Blinking of LED will be stopped by pressing the front panel key other than the
Open/Close, Fast, or arrow keys.
NOTE | When Kelvin cables are connected to Outputs |
| For the Kelvin outputs, the |
| • For the input couple port: |
| Control the switches to connect the input couple port to the Kelvin output. |
| • For the single input port: |
| Control the switches to connect the input port to two ports of the Kelvin output. |
| If you do not comply with this note, leakage current between the Kelvin paths will |
| cause measurement error. |
Agilent B2200 User’s Guide, Edition 2 |