Agilent Technologies B2201A manual DIAGnosticTESTFRAMeSTATe?, ROUTeAGNDCHANnelDISableCARD

Models: B2201A

1 234
Download 234 pages 49.02 Kb
Page 150
Image 150

SCPI Command Reference :DIAGnostic:TEST:FRAMe:STATe?

:DIAGnostic:TEST:FRAMe:STATe?

This command returns the most recent test result (pass/fail) of the specified test.

Syntax

:DIAGnostic:TEST:FRAMe:STATe? CONTroller FPANel LED

 

PEN BEEPer

 

 

Parameter

Explanation

 

 

 

 

CONTroller

Controller test

 

 

 

 

FPANel

Front panel interface test

 

 

 

 

LED

LED matrix test

 

 

 

 

PEN

Light pen test

 

 

 

 

BEEPer

Beeper test

 

 

 

Query response

test_result <newline><^END>

 

1: fail

 

 

0: pass

 

 

-1: not tested

 

Example

OUTPUT @Agb2200;":DIAG:TEST:FRAM:STAT? CONT"

 

ENTER @Agb2200;A

 

 

[:ROUTe]:AGND:CHANnel:DISable:CARD

 

This command ground-disables the all output ports (channels) for the specified card.

 

When the Ground Mode is ON, the ground-disabledoutput ports are disconnected

 

from the input Ground Port. At *RST, no channel is ground-enabled. The ground

 

mode is set by “[:ROUTe]:AGND[:STATe]”.

Syntax

[:ROUTe]:AGND:CHANnel:DISable:CARD card_number

 

 

 

 

Parameter

Explanation

 

 

 

 

card_number

For Auto Config mode: 0 or ALL

 

 

For Normal Config mode: 1, 2, 3, 4, or ALL

Example

OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS:CARD ALL"

5-24

Agilent B2200 User’s Guide, Edition 2

Page 150
Image 150
Agilent Technologies B2201A manual DIAGnosticTESTFRAMeSTATe?, ROUTeAGNDCHANnelDISableCARD