Programming Commands

C1 [one] Set A Minus B Mode Off

C1 [one]

Set A Minus B Mode Off

Syntax

C1;

LF

CR

SP

,

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

The C1 command turns the A Minus B mode off. That is, it switches off the functionality that was switched on by the C2 command (page 120) or by the AMB ON command (page 100).

NOTE

The functions of the command C1 are identical to the AMB OFF command

 

(page 100).

 

 