Programming Commands

CLRW Clear Write

CLRW

Clear Write

Syntax

CLRW SP

TRA ;

TRB

Preset State: CLRW TRA

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Clears the specified trace and enables trace data acquisition. The CLRW command places the indicated trace in clear-write mode. Data acquisition begins at the next sweep. (See the TS command (page 344) for more information about data acquisition.)

TRA corresponds to Trace 1 and TRB corresponds to Trace 2.

NOTE

The functions of the command CLRW are identical to the A1 command (page 79)

 

and B1 command (page 110).

On the 8560 series, 8566A/B, and 8568A/B analyzers the trace settings are set by the trace mode parameters, CLRW, VIEW, BLANK, MINH and MAXH and the averaging settings by VAVG. On the X-series the same settings are set in the Trace/Detector and View/Blank parameters. The following table describes what the N9061A application sets in the X-series box when the legacy commands for trace mode and averaging are sent.