Programming Commands

ML Mixer Level

NOTE

If an external amplifier gain value is set, the mixer level is determined using the

 

following equation:

 

Mixer Level = Ref. Level - Attenuation + Ext. Amplifier Gain

 

The external amplifier gain is not preset by doing an IP command in case the

 

analyzer is measuring a large signal. This is to protect the analyzer from damage

 

from a large signal. For a helpful suggestion, see Chapter 3 , “Hints and Tips,” on

 

page 71.

 

 

 

 

NOTE

For 8566A/B, 8568A/B, the functions of the ML command are identical to the KS,

 

command (page 179).