Agilent Technologies N9030a manual Mksp, Marker Span

Models: N9030a

1 368
Download 368 pages 42.13 Kb
Page 261
Image 261

Programming Commands

MKSP Marker Span

MKSP

Marker Span

Syntax

MKSP ;

LF

CR

SP

,

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

The MKSP command operates only when the delta marker is On (see MKD (page 245) or M3

(page 229)). When the delta marker is On and MKSP is executed, the delta marker and active marker determine the start and stop frequencies. The left marker specifies the start frequency, and the right marker specifies the stop frequency. If marker delta is Off, there is no operation.

NOTE

If the active marker is not a delta marker, there is no change in its position.

 

 

NOTE

For 8566A/B, 8568A/B, The functions of the MKSP command are identical to the

 

KSO command (page 207).

 

 

261