Programming Commands

LIMIFAIL Limits Failed

LIMIFAIL

Limits Failed

Syntax

LIMIFAIL ? ;

LF

CR

SP

,

Legacy Products

8560 series

Description

The LIMIFAIL command returns a number between 0 and 3 which specifies whether the active trace passed or failed the upper and lower limit line tests. The meanings of the returned numbers are shown below.

Table 4-23

Results of the LIMIFAIL Command

 

 

Result

Meaning

0The active trace passed both the upper and the lower limit tests.

1The active trace failed the lower limit test.

2The active trace failed the upper limit test.

3The active trace failed both the upper and the lower limit tests.