![](/images/new-backgrounds/1033321/33321219x1.webp)
Programming Commands
B1 [one] Clear Write for Trace B
B1 [one]
Clear Write for Trace B
Syntax
B1;
LF
CR
SP
,
Legacy Products
8560 series, 8566A/B, 8568A/B
Description
The B1 command sets Trace B to clear write. That is, it continuously displays any signal present at the analyzer input. The B1 command initially clears Trace B, setting all elements to zero. The sweep trigger then signals the start of the sweep, and Trace B is continually updated as the sweep progresses.
Subsequent sweeps send new amplitude information to the display addresses.
NOTE | The functions of the command B1 are identical to the CLRW TRB command |
| (page 130). |
110