Programming Commands

SWPCPL Sweep Couple

SWPCPL

Sweep Couple

Syntax

SWPCPL SP

SA ;

SRLF

?

CR

SP

,

The preset value is SA.

Legacy Products

8560 series

Description

Selects either a stimulus-response (SR) or signal-analyzer (SA) auto-coupled sweep time. In stimulus response mode, auto-coupled sweep times are usually much faster for swept response measurements. Stimulus response auto-coupled sweep times are typically valid in stimulus-response measurements when the system frequency span is less than 20 times the bandwidth of the device under test.