ROFFSET Reference Level Offset

ROFFSET

Reference Level Offset

Syntax

8560 series:

ROFFSET SP

SP

?

real

UP

DN

DB

;

LF

CR

SP

,

8566A/B, 8568A/B:

ROFFSET SP

?

real

OA

DM ;

MV

UV

DB

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Offsets all amplitude readouts without affecting the trace.

Once activated, the ROFFSET command displays the amplitude offset on the left side of the screen. Entering ROFFSET 0 or presetting the analyzer eliminates an amplitude offset.

NOTE

For 8566A/B, 8568A/B, the functions of the ROFFSET command are identical to

 

the KSZ command (page 214).

 

 