KS( Lock Registers

KS(

Lock Registers

Syntax

KS( ;

Legacy Products

8566A/B, 8568A/B

Description

The KS( command secures the contents of registers one through six. When the registers are secured, the SV and SAVE commands cannot save more instrument states in the registers, but instead write "SAVE LOCK" on the analyzer display. To save an instrument state in a locked register, first execute KS) to unlock the registers.

The recall function of the analyzer is not affected by this function.