KSl Transfer Trace B to Trace C

KSl

Transfer Trace B to Trace C

Syntax

KSl;

Legacy Products

8566A/B, 8568A/B

Description

The KSl command transfers Trace B data to Trace C

NOTE

Trace C cannot be an active trace. This means that the data in Trace C cannot be

 

updated as the analyzer sweeps. To ensure that the current settings of the analyzer

 

are reflected in the data transferred from Trace B to Trace C, you must follow the

 

four step process below.

Select single sweep mode (S2 (page 307) or SNGLS command (page 313))

Select the desired analyzer settings

Take one complete sweep using the TS command (page 344)

Transfer the data

NOTE

The functions of the command KSl are identical to the BTC command (page 117).