![](/images/new-backgrounds/1033321/33321207x1.webp)
Programming Commands
AT Input Attenuation
AT
Input Attenuation
Syntax
8560 Series:
AT
SP number
SP
UP
DN
AUTO
MAN
OA
?
DB ;
LF
CR
SP
,
8566A/B, 8568A/B:
AT | SP | number |
SP
UP
DN
OA
?
Table
DB ;
Item | Description/Default | Range |
|
|
|
Number | Any real number or | 0 to 70 dB specified absolutely |
| integer. If the value you | and |
| enter is not a valid value | 10 to 70 dB in 10 dB steps1 |
| for the analyzer you are |
|
| using, it switches |
|
| automatically to the |
|
| closest valid setting. |
|
| Default units are dB. |
|
|
|
|
1. The range is limited to 0 to 60 dB if 8564E/EC or 8565E/EC is selected.
Preset State: 10 dB
Step Increment: 10 dB
104