Programming Commands

 

 

 

ERR Error

Table 4-7

 

 

 

 

 

 

 

 

 

X-series

Description

8560 series

Description

 

error code

 

error code

 

 

 

 

 

 

 

40

TG Alignment Failure

758

SYSTEM: Unknown system error

 

 

 

 

 

 

42

RF Alignment Failure

758

SYSTEM: Unknown system error

 

 

 

 

 

 

44

IF Alignment Failure

758

SYSTEM: Unknown system error

 

 

 

 

 

 

46

LO Alignment Failure

758

SYSTEM: Unknown system error

 

 

 

 

 

 

48

ADC Alignment Failure

758

SYSTEM: Unknown system error

 

 

 

 

 

 

50

FM Demod Alignment

758

SYSTEM: Unknown system error

 

 

Failure

 

 

 

 

 

 

 

 

54

Extended Align Failure

758

SYSTEM: Unknown system error

 

 

Sum

 

 

 

 

 

 

 

 

71

Characterize Preselector

758

SYSTEM: Unknown system error

 

 

Failure

 

 

 

 

 

 

 

 

-200.3310

Execution Error;

758

SYSTEM: Unknown system error

 

 

Preselector Centering

 

 

 

 

failed

 

 

 

 

 

 

 

 

503

Frequency Reference

336

10 MHz Ref Cal oscillator failed to lock

 

 

Unlocked

 

when going to internal 10 MHz

 

 

 

 

reference.

 

 

 

 

 

 

505

2nd LO Unlocked

336

10 MHz Ref Cal oscillator failed to lock

 

 

 

 

when going to internal 10 MHz

 

 

 

 

reference.

 

 

 

 

 

 

509

LO Unlocked

300

YTO UNL: YTO (1st LO) phase-locked

 

 

 

 

loop (PLL) is unlocked.

 

 

 

 

 

 

513

IF Synthesizer Unlocked

450

IF SYSTM: IF hardware failure. Check

 

 

 

 

other error messages.

 

 

 

 

 

 

515

Calibration Oscillator

336

10 MHz Ref: Cal oscillator failed to lock

 

 

Unlocked

 

when going to internal 10 MHz

 

 

 

 

reference

 

 

 

 

 

 

521

External Ref missing or

905,333

EXT REF: Unable to lock cal oscillator

 

 

out of range

 

when set to external reference. Check

 

 

 

 

that the external reference is within

 

 

 

 

tolerance.

 

 

 

 

600 UNLK: 600 MHz reference

 

 

 

 

oscillator PLL is unlocked

 

 

 

 

 

 

153

Page 153
Image 153
Agilent Technologies N9030a manual 153

N9030a specifications

Agilent Technologies, now known as Keysight Technologies, is recognized as a leader in providing advanced electronic measurement solutions. One of their highlighted products is the N9030A PXA X-Series Signal Analyser, a versatile instrument that caters to the needs of engineers and researchers in various fields, including telecommunications, aerospace, and defense.

The Agilent N9030A is designed to address the increasing complexity of modern signals. One of its standout features is its exceptional frequency range, spanning from 3 Hz to 50 GHz, making it suitable for both wideband and narrowband applications. This broad frequency coverage allows users to capture a wide array of signals, from low-frequency analog to high-frequency digital communications.

Another significant characteristic of the N9030A is its high dynamic range, which enables accurate measurements even in the presence of strong interfering signals. This capability is essential for testing and developing communication systems that operate in challenging environments, where signal integrity is paramount. The analyser also features a high sample rate of 5 GS/s, providing detailed resolution for even the most intricate signals.

The integrated measurement applications further enhance the instrument’s versatility, allowing for specific analyses related to modulation formats, spectral emissions, and time-domain events. Such applications streamline workflows and improve measurement efficiency, making it easier for users to obtain and interpret data.

One of the technological highlights of the N9030A is its advanced digital signal processing (DSP), which significantly improves measurement speed and performance. The DSP capabilities allow engineers to conduct real-time analysis, which is critical for applications that require immediate feedback, such as signal troubleshooting and validation.

The user-friendly interface of the N9030A, along with a myriad of connectivity options, ensures that integration into existing testing setups is seamless. Data can be easily exported to other software tools for further analysis, enhancing collaboration and report generation.

In conclusion, the Agilent N9030A PXA X-Series Signal Analyser stands out as a powerful tool for modern signal analysis. With its wide frequency range, high dynamic range, advanced DSP capabilities, and integrated measurement applications, it is tailored to meet the rigorous demands of today’s engineers and researchers. Whether for development, testing, or troubleshooting, the N9030A provides the performance and flexibility required for successful project outcomes.