Programming Commands

CR Couple Resolution Bandwidth

CR

Couple Resolution Bandwidth

Syntax

CR;

LF

CR

SP

,

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

The CR command couples the resolution bandwidth to the span.

The counterpart to the CR command is the RB command (page 295) which breaks the coupling. Use the CR command to re-establish coupling after executing an RB command.

NOTE

CR uses the legacy instrument settings for resolution bandwidth only if Mode

 

Setup > Preferences> Limit RBW/VBW is set to ON.

 

 