Agilent Technologies N9030a manual Rcls

Models: N9030a

1 368
Download 368 pages 42.13 Kb
Page 299
Image 299

Programming Commands

RCLS Recall State

RCLS

Recall State

Syntax

RCLS SP

SP

LAST;

PWRON

LF

number

CR

 

 

SP

,

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Recalls analyzer state data from the specified state register in the analyzer’s memory.

Registers one through six are reserved for the user, and contain instrument states (such as front panel configuration) saved with the SAVES command (page 308) or the SV command (page 324).

NOTE

The functions of the RCLS command are identical to the RC command (page 298).

299

Page 299
Image 299
Agilent Technologies N9030a manual Rcls