Programming Commands

KS, Mixer Level

KS,

Mixer Level

Syntax

KS, SP

?

real

OA

DM ;

MV

UV

Legacy Products

8566A/B, 8568A/B

Description

The KS, command specifies the maximum signal level that is applied to the input mixer for a signal that is equal to or below the reference level.

The effective mixer level is equal to the reference level minus the input attenuator setting. When KS, is activated, the effective mixer level can be set from -10 dBm to -70 dBm in 10 dB steps.

As the reference level is changed, the coupled input attenuator automatically changes to limit the maximum signal at the mixer input to your specified setting for signals less than or equal to the reference level.

NOTE

If the external amplifier gain has been set, executing an IP command will not reset

 

this value. This is to protect the analyzer.

 

 

NOTE

The functions of the KS, command are identical to the ML command (page 267).

 

 