Programming Commands

AT Input Attenuation

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

Specifies the RF input attenuation.

Although the attenuation level in the X-series of analyzers can be specified using absolute values, you can never set attenuation below 10 dB using the DOWN steps. This is a safety feature to prevent inadvertent setting of attenuation to a level that could damage the analyzer.

NOTE

In PXA/MXA, the auto attenuation range is 6-70 dB and in EXA, it is 6-60 dB.

 

 

CAUTION

Signal levels above +30 dBm will damage the signal analyzer.

 

 

 

 

NOTE

You cannot step down below 10 dB. To set levels below 10 dB, you must specify

 

the attenuation absolutely. For example, to set attenuation to 0 dB, you must use

 

the command AT 0DB.

 

 