Agilent Technologies N9030a manual Start Frequency, Khz Mhz Ghz Kz Sp

Models: N9030a

1 368
Download 368 pages 42.13 Kb
Page 156
Image 156

Programming Commands

FA Start Frequency

FA

Start Frequency

Syntax

FA SP

SP

?

Table 4-8

real

UP

DN

OA

HZ ;

KHZ

LF

MHZ

CR

GHZ

KZ SP

MZ,

GZ

Item

Description/Default

Range

 

 

 

REAL

Any real or integer number.

Frequency range of the analyzer

 

Default unit is Hz.

 

 

 

 

Step Increment: Frequency span divided by 10

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

The FA command specifies the start frequency value. The start frequency is equal to the center frequency minus (the span divided by two) (FA = CF − SP/2). Changing the start frequency changes the center frequency and span.

NOTE

The OA parameter only returns the current value to the controller. It does not set

 

the active function to the start frequency.

 

 

156

Page 156
Image 156
Agilent Technologies N9030a manual Start Frequency, Khz Mhz Ghz Kz Sp