Programming Commands

OL Output Learn String

OL

Output Learn String

Syntax

OL

SP

80 byte binary

string

 

 

?

;

Legacy Products

8566A/B, 8568A/B

Description

The OL command transmits information to the controller that describes the state of the analyzer when the OL command is executed. This information is called the “learn string.” The learn string can be sent from the controller memory back to the analyzer to restore the analyzer to its original state.

NOTE

The OL command is not completely supported, due to differences between the

 

X-Series and 8566/8568. The following table outlines each byte of the array and

 

the bits supported within that byte.

 

 

Byte

Support Information

 

 

1

Fixed decimal value 31

 

 

2

Fixed decimal value 118

 

 

3 to 9

Supported

 

 

10

Supported

 

 

11

Unsupported: Fixed decimal 0

 

 

12 to 17

Supported

 

 

18

Supported: Bits 6, 2, 1 and 0

 

Unsupported: Bits 7, 5, 4, and 3

 

 