Agilent Technologies N9030a manual Fdsp, Frequency Display Off

Models: N9030a

1 368
Download 368 pages 42.13 Kb
Page 158
Image 158

Programming Commands

FDSP Frequency Display Off

FDSP

Frequency Display Off

Syntax

FDSP SP OFF ;

?LF

CR

SP

,

Legacy Products

8560 series

Description

The FDSP command turns the frequency annotation OFF.

NOTE

It is not possible enable or disable the frequency annotation alone, leaving other

 

annotation unaffected. Thus, the FDSP command behaves in the same way as

 

ANNOT (page 102) If the FDSP command has been used to disable the frequency

 

annotation, sending the ANNOT ON command does not re-enable the display

 

annotation. The display annotation is only displayed by sending the IP (page 178)

 

command.

 

 