Programming Commands

M2 [two] Marker Normal

M2 [two]

Marker Normal

Syntax

M2

SP

real

SP UP

DN

OA

?

HZ ;

KHZLF

MHZ

CR

GHZ

KZ SP

MZ,

GZ

S

MS

US

SC

Legacy Products

8566A/B, 8568A/B

Description

The M2 [two] command moves the active marker to the marker frequency. If the active marker type is not currently normal (for example, if it is delta), the M2 command changes it to a normal marker.

NOTE

The functions of the M2 command are identical to the MKN command (page 251).

 

 

NOTE

If the active marker has not been declared with MKACT, a normal marker is

 

turned on and this active marker is assumed to be marker number 1 [one].