Agilent Technologies N9030a manual Bml, Trace B Minus Display Line

Models: N9030a

1 368
Download 368 pages 42.13 Kb
Page 116
Image 116

Programming Commands

BML Trace B Minus Display Line

BML

Trace B Minus Display Line

Syntax

BML;

LF

CR

SP

,

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

The BML command subtracts the display line from trace B (point by point), and sends the difference to trace B. Trace B corresponds to Trace 2.

The command BML is calculated differently depending on the language being used; for the 8560 Series the calculation uses units of dBm.

For the 8566A, HP8566B, HP8568A, HP8568B - the BML command uses display units.

NOTE

The functions of the command BML are identical to the BL command (page 114).

116

Page 116
Image 116
Agilent Technologies N9030a manual Bml, Trace B Minus Display Line