Programming Commands

FB Stop Frequency

FB

Stop Frequency

Syntax

FB SP

SP

?

Table 4-9

real

UP

DN

OA

HZ ;

KHZ

LF

MHZ

CR

GHZ

KZ SP

MZ,

GZ

Item

Description/Default

Range

 

 

 

REAL

Any real or integer number.

Frequency range of the analyzer

 

Default unit is Hz.

 

 

 

 

Step Increment: Frequency span divided by 10

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

The FB command specifies the stop frequency value. The stop frequency is equal to the center frequency plus the span divided by two (FB = CF + SP/2). Changing the stop frequency changes the center frequency and span.

NOTE

The OA parameter only returns the current value to the controller. It does not set

 

the active function to the stop frequency.

 

 