Agilent Technologies N9030a manual M3 three Delta Marker

Models: N9030a

1 368
Download 368 pages 42.13 Kb
Page 229
Image 229

Programming Commands

M3 [three] Delta Marker

M3 [three]

Delta Marker

Syntax

M3

SP

real

SP UP

DN

OA

?

HZ ;

KHZLF

MHZ

CR

GHZ

KZ SP

MZ,

GZ

S

MS

US

SC

Legacy Products

8566A/B, 8568A/B

Description

The M3 [three] command computes the frequency and amplitude difference between the active marker and the delta (or difference) marker.

If a delta marker is not displayed on the screen, the M3 command places one at the specified frequency or on the right hand edge of the display. If an active marker is not displayed on the screen, the M3 command places an active marker at the center of the screen.

NOTE

The active marker is the number 1 marker unless otherwise specified by the

 

MKACT command (page 242).

 

 

 

 

NOTE

The functions of the M3 command are identical to the MKD command (page 245).

 

 