Agilent Technologies N9030a manual Preselector Peak

Models: N9030a

1 368
Download 368 pages 42.13 Kb
Page 288
Image 288

Programming Commands

PP Preselector Peak

PP

Preselector Peak

Syntax

PP

;

LF

CR

SP

,

Legacy Products

8560 series, 8566A/B

Description

The PP command optimizes preselector tracking to peak the amplitude of a signal at the active marker. If a marker is not on the screen, PP places a marker at the highest signal level, and optimizes preselector tracking at that frequency.

NOTE

This command is only supported when the X-series analyzer’s maximum

 

frequency limit is greater than 3.6 GHz. If the command is issued on an analyzer

 

with a maximum frequency limit of 3.6 GHz or less the command is not

 

implemented and no error is generated.