The $prompt will appear on the screen when
the ALC Bandwidth calibration is complete.
(This can take up to 15 minutes depending on
the frequency range of the 682XXB/683XXB.)
b. Record step completion on the Test Record.
2. Store the calibration data as follows:
a. At the $prompt, type: calterm 787 and
press <ENTER>. (The $prompt will appear on
the screen when the calibration data has been
stored.)
b. Record step completion on the Test Record.
4-32 682XXB/683XXB MM
CAUTION
Whensaving calibration data, turn-
ing off the instrument before the $
prompt returns to the screen can
cause all stored data to be lost.
ALC BANDWIDTH
CALIBRATION CALIBRATION