5. Perform AM Function Generator calibration as
follows:
a. At the $prompt, type: calterm 146 and
press <ENTER>. (The $prompt will appear on
the screen when the calibration is complete.)
b. Record step completion on the Test Record.
6. Store the calibration data as follows:
a. At the $prompt, type: calterm 787 and
press <ENTER>. (The $prompt will appear on
the screen when the calibration data has been
stored.)
b. Record step completion on the Test Record.
4-36 682XXB/683XXB MM
AM
CALIBRATION CALIBRATION
CAUTION
Whensaving calibration data, turn-
ing off the instrument before the $
prompt returns to the screen can
cause all stored data to be lost.