682XXB/683XXB MM A-103
TEST MODEL
RECORD 68259B/68359B
3-11 Power Level Accuracy and Flatness Tests (Continued)

(Model 68259B/68359B with Option 15B High Power & Option 2A Step Attenuator)

Power Level Accuracy Test Procedure
Set F1 to 1.0 GHz Set F1 to 5.0 GHz Set F1 to 22.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power * Set L1 to: Measured Power *
+11dBm
+10dBm
+9 dBm
+8 dBm
+7 dBm
+6 dBm
+5 dBm
+4 dBm
+3 dBm
+2 dBm
+1 dBm
+0 dBm
–1 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+11dBm
+10dBm
+9 dBm
+8 dBm
+7 dBm
+6 dBm
+5 dBm
+4 dBm
+3 dBm
+2 dBm
+1 dBm
+0 dBm
–1 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+7.5 dBm
+6.5 dBm
+5.5 dBm
+4.5 dBm
+3.5 dBm
+2.5 dBm
+1.5 dBm
+0.5 dBm
–0.5 dBm
–1.5 dBm
–2.5 dBm
–3.5 dBm
–4.5 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
*Specification is 1.0dB. *Specification is 1.0dB. *Specification is 1.0dB.
Power Level Flatness Test Procedure (Step Sweep)
Set L1 to: Max Power Min Power Variation **
+7.5 dBm _________dBm __________dBm _________dB
**Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 26.5 GHz).
Power Level Flatness Test Procedure (Analog Sweep) (Model 68359B only)
Set L1 to: Max Power Min Power Variation ***
+7.5 dBm _________dBm __________dBm _________dB
***Maximum variation is 7.0 dB (0.01 to 0.05 GHz); 6.0 dB (0.05 to 20 GHz); 8.2 dB (20 to 26.5 GHz)(typical, not a
specification).