682XXB/683XXB MM A-59
TEST MODEL
RECORD 68253B/68353B
3-9 Harmonic Test: RF Output Signals From 2 to 20 GHz (Continued)
Test Procedure (11 to 20 GHz) Measured Value Upper Limit
SetF1 to 12.4 GHz
Recordthe level of all harmonics of the 12.4 GHz carrier:
24.8GHz (2nd harmonic) .....................
37.2GHz (3rd harmonic) ......................
_______________dBc
_______________dBc
–60dBc*
–60dBc*
SetF1 to 16.0 GHz
Recordthe level of all harmonics of the 16.0 GHz carrier:
32.0GHz (2nd harmonic) ..................... _______________dBc –60dBc*
SetF1 to 20.0 GHz
Recordthe level of all harmonics of the 20.0 GHz carrier:
40.0GHz (2nd harmonic) ..................... _______________dBc –60dBc*
*–50 dBc is Option 15B (High Power) installed.