682XXB/683XXB MM 5-23±15VLP Supply Problems
This supply provides ±15 volts to the Pulse Generator and the Refer-
ence, Coarse, Fine, and YIG Loop circuits.
Step 6. Perform the following procedure to isolate malfunctions to
the ±15VLP supply and outlying load circuits.
a. Place the LINE switch to STANDBY.
b. Replace the A15 PCB assembly.
c. Place the LINE switch to OPERATE and measure the
±15VLP voltages per Table5-7.
If the voltages are correct, the problem is cleared.
If the voltages are incorrect, go to step d.
d. Place the LINE switch to STANDBY.
e. Remove the A3, A4, A5, A6, and A7 PCBs.
f. Place the LINE switch to OPERATE and measure the
±15VLP voltages.
If the voltages are correct, go to step g.
If the voltages are still incorrect, contact your local
ANRITSU service center for assistance.
g. Place the LINE switch to STANDBY,then install one of
the removed PCBs.
h. Placethe LINE switch to OPERATE and measure the
±15VLP voltages.
i. Continue steps g and h until the faulty PCB is located.
Table5-6. Error Messages 105 and 106 (5 of 10)
TROUBLESHOOTING
TROUBLESHOOTING TABLES