3-4 682XXB/683XXB MM

INSTRUMENT CRITCAL
SPECIFICATION RECOMMENDED
MANUFACTURER/MODEL TEST
NUMBER
SpectrumAnalyzer,
with
ExternalMixers
and
DiplexerAssy
FrequencyResolution: 0.01 to 65 GHz
ResolutionBandwidth: 10 Hz Tektronix,Model 2794, with
ExternalMixers:
WM780K(18 to 26.5 GHz)
WM780A(26.5 to 40 GHz)
WM780U(40 to 60 GHz)
WM780E(60 to 90 GHz)
and
DiplexerAssy: 015-0385-00
3-8,3-9, 3-12,
3-13,3-16,
3-17
SpectrumAnalyzer FrequencyRange: 20 Hz to 40 MHz
ResolutionBandwidth: 3Hz Hewlett-Packard,Model 3585B 3-10
FrequencyCounter
with
CableKit
and
ExternalMixer
FrequencyRange: 0.01 to 65 GHz
InputImpedance: 50
Resolution:1Hz
Other:External Time Base Input
EIPMicrowave, Inc. Models 538B,
548B,or 578B, with
CableKit: Option 590
and
ExternalMixer:
Option91 (26.5 to 40 GHz)
Option92 (40 to 60 GHz)
Option93 (60 to 90 GHz)
3-7
PowerMeter,
with
PowerSensor
PowerRange: –30 to +20 dBm
(1 Wto 100mW) Hewlett-PackardModel 437B, with
PowerSensor:
HP8487A (0.01 to 50 GHz)
3-11
PowerMeter
with
PowerSensor
PowerRange: –30 to +20 dBm
(1 Wto 100mW) ANRITSUML4803A, with
PowerSensor:
MP716A4(50 to 75 GHz)
3-11
FrequencyReference Frequency:10 MHz
Accuracy:5x10
–12parts/day Absolute Time Corp., Model 300 3-6
ModulationAnalyzer FrequencyInput: 10 MHz
(orthe IF of the spectrum analyzer)
AMDepth: 0% to 90%
AMModulation Rates: dc to 100 kHz
Filters:20 kHz lowpass, 300 Hz highpass
Hewlett-Packard,Model 8901A 3-12
Oscilloscope Bandwidth:DC to 150 MHz
VerticalSensitivity: 2mV/division
HorizontalSensitivity: 50ns/division
Tektronix,Inc. Model TAS485 3-6, 3-11,
3-14,3-15
Mixer FrequencyRange: 1 to 26 GHz MacomMicro Electronics Div.
ModelDMS1-26A 3-10
Adapter K (male) to 2.4 mm (female)
Adaptsthe Power Sensor, HP 8487A, to
the682XXB/683XXB RF OUTPUT con-
nector( 40 GHz models)
Hewlett-Packard
PartNumber: HP 11904D 3-11
Adapter Adapts the MP716A4 Power Sensor to
theML4803A Power Meter ANRITSU MA4002B 3-11
Table3-1. Recommended TestEquipment for Performance Verification Tests(1 of 2)
PERFORMANCE RECOMMENDEDVERIFICATION TEST EQUIPMENT