ERROR AND WARNING/

TROUBLESHOOTING

STATUS MESSAGES

 

Table 5-3.Possible Warning/Status Messages during Normal Operation

 

 

 

 

 

 

Warning/Status

Description

 

 

Message

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This warning message indicates that the 100 MHz

 

 

OVN COLD

 

 

 

 

 

 

 

 

 

 

 

 

Crystal oven (or the 10 MHz Crystal oven if Option 16 is

 

 

 

 

 

 

 

 

 

 

 

 

installed) has not reached operating temperature.

 

 

 

 

 

 

 

 

 

 

 

 

Normally displayed during a cold start of the signal

 

 

 

 

 

 

 

 

 

 

 

 

generator. If the message is displayed during normal

 

 

 

 

 

 

 

 

 

 

 

 

operation, it could indicate a malfunction. Run self-test

 

 

 

 

 

 

 

 

 

 

 

 

to verify.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Displayed when the RF output goes unleveled. Normally

 

 

UNLEVELED

 

 

 

 

 

 

 

 

 

 

 

 

caused by exceeding the specified leveled-power rating.

 

 

 

 

 

 

 

 

 

 

 

 

Reducing the power level usually clears the warning

 

 

 

 

 

 

 

 

 

 

 

 

message.

 

 

 

 

 

 

 

 

 

 

 

 

If the warning message is displayed only when AM is

 

 

 

 

 

 

 

 

 

 

 

 

selected ON, the modulating signal may be driving the

 

 

 

 

 

 

 

 

 

 

 

 

RF output unleveled. Reducing the modulating signal or

 

 

 

 

 

 

 

 

 

 

 

 

adjusting the power level usually clears the warning.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

When Unlocked/Narrow FM or Unlocked/Wide FM is

 

 

UNLOCKED

 

 

 

 

 

 

 

 

 

 

 

 

selected ON, this warning message appears indicating

 

 

 

 

 

 

 

 

 

 

 

 

that the instrument is not phase-locked during this FM

 

 

 

 

 

 

 

 

 

 

 

 

mode of operation.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This warning message is displayed when the AM Rate

 

 

Reduce Rate

 

 

 

 

 

 

 

 

 

 

 

 

or FM Rate is set >100 kHz for a non-sinewave

 

 

 

 

 

 

 

 

 

 

 

 

modulating waveform. Amplitude or frequency

 

 

 

 

 

 

 

 

 

 

 

 

modulation of the output signal will continue but the

 

 

 

 

 

 

 

 

 

 

 

 

modulating waveform may be distorted.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This status message indicates that an external 10 MHz

 

 

EXT REF

 

 

 

 

 

 

 

 

 

 

 

 

signal is being used as the reference signal for the

 

 

 

 

 

 

 

 

 

 

 

 

682XXB/683XXB.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This status message indicates that a constant (offset)

 

 

OFFSET

 

 

 

 

 

 

 

 

 

 

 

 

has been applied to the displayed power level.

 

 

 

 

 

 

 

 

 

 

 

 

 

This status message indicates that a power slope

 

 

SLOPE

 

 

 

 

 

 

 

 

 

 

 

 

correction has been applied to the ALC.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This status message indicates that a user level flatness

 

 

USER 1...5

 

 

 

 

 

 

 

 

 

 

 

 

correction power-offset table has been applied to the

 

 

 

 

 

 

 

 

 

 

 

 

ALC.

 

 

 

 

 

 

 

 

 

 

 

 

This status message indicates that the 682XXB/683XXB

 

 

SS MODE

 

 

 

 

 

 

 

 

 

 

 

 

has been placed in a source lock mode for operation

 

 

 

 

 

 

 

 

 

 

 

 

with a 360B Vector Network Analyzer.

 

 

 

 

 

 

 

 

 

 

 

 

 

682XXB/683XXB MM

5-9

Page 175
Image 175
Anritsu 683XXB, 682XXB manual OVN Cold, Message

682XXB, 683XXB specifications

The Anritsu 683XXB and 682XXB series are advanced vector network analyzers (VNAs) renowned for their precision and versatility in characterizing RF and microwave components. Designed for engineers and technicians involved in the development, manufacturing, and testing of high-frequency devices, these analyzers offer state-of-the-art technology that ensures optimal performance in various applications.

One of the hallmark features of the Anritsu 683XXB and 682XXB is their high dynamic range, which allows for accurate measurements of small reflection and transmission coefficients, essential for assessing the performance of complex RF structures. With frequency coverage extending from DC to 70 GHz, these analyzers cater to a broad spectrum of applications, making them suitable for industries such as telecommunications, aerospace, and automotive.

The user-friendly interface of the Anritsu VNAs is complemented by a high-resolution display, which facilitates easy navigation through measurement setups and results. The analyzers feature multiple measurement modes, including S-parameter measurements, time-domain analysis, and noise figure measurements, providing engineers with comprehensive tools for device characterization.

Both the 683XXB and 682XXB implement advanced calibration techniques, including automated calibration and error correction methods, to enhance measurement accuracy. These methods significantly reduce the uncertainties associated with test setups, enabling reliable performance evaluations of components like filters, amplifiers, and antennas.

Anritsu’s proprietary technologies, such as the VectorStar and ShockLine series integration, further empower the 683XXB and 682XXB models. These technologies enable high-throughput testing and improved measurement stability, addressing the needs of modern production environments that demand rapid turnaround times without sacrificing precision.

Additionally, the analyzers come equipped with various connectivity options, including USB, LAN, and GPIB, ensuring seamless integration into automated test systems. This adaptability enhances the analyzers' utility in both laboratory settings and field operations.

In conclusion, the Anritsu 683XXB and 682XXB series vector network analyzers represent the pinnacle of RF and microwave testing technology. With their unmatched precision, comprehensive measurement capabilities, and advanced calibration techniques, these instruments are indispensable tools for professionals striving to push the boundaries of high-frequency device performance and reliability.